UV spot curing
OmniCure S2000 / S2000-XLAConsumables & service life
The S2000 and S2000-XLA platform, still in service across the region. Its lamp module and filters are not the Elite's: check the part number before ordering, because the two sets look alike and will not fit each other. Which filter is fitted to your machine is ticked on the label on its back panel β read that before ordering a replacement.
About the system itselfLamps
S2000 replacement lamp
The 200 W replacement lamp for the S2000 and S2000-XLA platform, Standard and Surface Cure. It carries Intelli-Lamp too, so the system reads the hours off the lamp itself. Same 200 W mercury arc source as the Elite's; the module differs and the two are not interchangeable.
Product detailsNever interrupt the warm-up cycle: a lamp shut down mid-warm-up is not covered by the manufacturer's guarantee. The lamp contains mercury and its box says so β it is hazardous waste, and it must be disposed of under the law of the country it is in, never in general waste.
Service life
- Guaranteed
- 2,000 h
- Counted by
- Intelli-Lamp chip on the lamp, read by the system
Source: Manufacturer lamp guarantee
Replace it when
- Lamp-hour counter past the guaranteed figure and output falling
- Measured irradiance at the light guide tip has dropped against the last reading
- Cure time has crept up on an unchanged process
Part numbers carried by ETIA
012-64000RS2000 Replacement Lamp β Standard (200 W)012-65000RS2000 Replacement Lamp β Surface Cure (200 W)
These part numbers are on ETIA's item master β we buy and sell them in-region. Availability on the day is confirmed on the quotation.
Optical filters
S2000 / S1500 filter
Genuine Excelitas internal filters for the S2000, S2000-XLA and S1500, in five bands, plus external filters that mount ahead of the light guide. Which band is fitted is ticked on the label on the back of the machine.
Service life
No rated life is published. Same optical glass in the same hot beam path as the Elite cartridges: judge on measured output and on the glass.
Replace it when
- Measured irradiance has fallen although the lamp is new or its hours are low
- Visible clouding, browning or a burn mark on the glass
Part numbers carried by ETIA
019-01046RS2000 / S1500 Filter 250β450 nm019-01048RS2000 / S1500 Filter 320β390 nm019-01049RS2000 / S1500 Filter 320β500 nm019-01047RS2000 / S1500 Filter 400β500 nm019-01045RS2000 / S1500 Filter 365 nm019-01023320β390 nm External Filter019-01024365 nm External Filter019-01022320β480 nm External Filter019-01026400β500 nm External Filter019-01025Blank External Filter019-00108External Filter Adapter Kit
These part numbers are on ETIA's item master β we buy and sell them in-region. Availability on the day is confirmed on the quotation.
Light guides
Light guide
ETIA's own light guide line: liquid-filled and high-power fibre, 3, 5 and 8 mm tips, 750 to 5000 mm, single through four-leg, plus line terminations. This is the part that gets handled every shift, so it is the part that wears first.
Do not bend a light guide. Not into a curve, not into a coil, not by hanging it from its tip. There is no safe bend radius to work to β ETIA's rule is that the guide is not bent at all. Bending damage is permanent and is usually invisible from the outside: the first sign is a cure that will not pass. Route it straight, support its weight along its whole length, and cap the tip when it is off the machine.
Part numbers carried by ETIA
805-00002Liquid Light Guide (5 mm x 1000 mm)805-00004Liquid Light Guide (3 mm x 1000 mm)805-00007Liquid Light Guide (5 mm x 1500 mm)805-00009Liquid Light Guide (8 mm x 1000 mm)805-00011Liquid Light Guide (2 x 3 mm x 1000 mm)806-00005High-Power Fiber Light Guide (2 x 3 mm x 1000 mm)806-00011High-Power Fiber Light Guide (5 mm x 1000 mm)806-00012High-Power Fiber Light Guide (3 mm x 1000 mm)
These part numbers are on ETIA's item master β we buy and sell them in-region. Availability on the day is confirmed on the quotation.
Lenses & adapters
Light guide adapters & cure rings
What goes on the end of the guide to shape the beam: 90Β° angle adapters, adjustable collimating adapters, cure rings for 360Β° curing, light line assemblies.
Service life
These are glass and metal with no rated life. They come off the replacement list when the optic is scorched, fogged or chipped, or when a thread or thumb screw no longer holds position.
Replace it when
- Scorched or fogged lens face
- Chipped glass or a cracked holder
- Thumb screw or thread will not hold the set position
Part numbers carried by ETIA
810-00001Lightguide angle adapter, 90Β° for 3 mm LG810-00002Lightguide angle adapter, 90Β° for 5 mm LG810-00003Lightguide angle adapter, 90Β° for 8 mm LG810-000413 mm Adjustable Collimating Adapter (holder, lens, 2 thumb screws)810-000425 mm Adjustable Collimating Adapter (holder, lens, 2 thumb screws)810-000438 mm Adjustable Collimating Adapter (holder, lens, 2 thumb screws)810-00048Cure ring, slotted, 63.5 mm OD, 10 mm ID for 5 mm LG810-00049Cure ring, no-slot, 63.5 mm OD, 10 mm ID for 8 mm LG810-00050Cure ring, no-slot, 63.5 mm OD, 15 mm ID for 5 mm LG810-00013Light line assembly (50 mm Γ 0.1 mm)809-00083Quartz Rod Lens 3 mm (replacement for 810-00044)
These part numbers are on ETIA's item master β we buy and sell them in-region. Availability on the day is confirmed on the quotation.
Calibration & metrology
R2000 radiometer calibration
The radiometer is what makes every other figure on this page mean something. Its own calibration is consumed by use: the reference is valid for a set amount of lamp-on time, after which the readings are indications, not measurements.
Product detailsService life
- Calibration validity
- 112 h of lamp-on time
- Then
- Recalibrate before the readings are used for process control
Source: OmniCure S2000 Elite User Guide, 035-00707
Replace it when
- Lamp-on time since the last calibration has passed 112 h
- Two radiometers disagree on the same cure site
- An audit or a customer qualification needs a current certificate
Part numbers carried by ETIA
These part numbers are on ETIA's item master β we buy and sell them in-region. Availability on the day is confirmed on the quotation.
Safety & cleaning
Cleaning kit, end caps & safety glasses
The parts a bench actually gets through: the light guide cleaning kit, the coloured end caps that stop a tip being chipped on the bench, and UV safety glasses for everyone standing near an open beam.
Service life
Ordered by consumption, not by interval. The one item with a real replacement rule is the safety glasses: replace them when the lens is scratched, crazed or pitted, because scratched UV eyewear does not protect reliably.
Replace it when
- Safety glasses scratched, crazed or pitted β replace, do not polish
- Cleaning tissues or swabs run out mid-shift
- An end cap lost, so a tip is sitting bare on the bench
Part numbers carried by ETIA
019-00052Light guide cleaning kit (100 lens tissues, 20 pre-saturated tissues, 20 foam swabs, 250 mL wash bottle)019-00090RProtective End Cap, 3 mm (Red)019-00091RProtective End Cap, 5 mm (Blue)019-00092RProtective End Cap, 8 mm (Green)019-01010Light guide locking kit (for dual liquid light guides)854-00001RSafety glasses, UV protection grey, CE rated014-00070RFoot Switch (audio jack)
These part numbers are on ETIA's item master β we buy and sell them in-region. Availability on the day is confirmed on the quotation.
Order by part number, or ask first
Add the part numbers you need to the inquiry list and send them in one message. If you are not sure which part has gone, tell us the model, the lamp hours and what the cure is doing β we will work it out before you buy anything.